Configuring the MLC 8000 for IP-Based Simulcast
Prerequisites:
This process has the following prerequisites:
NoteWhen the MLC 8000 is configured for IP Simulcast, the base radios have to be the same type.
- Obtain the following:
- Radio Service Software (RSS) or Configuration/Service Software (CSS)
- Test Equipment (see Test Equipment)
- The following tasks must be completed:
- “Performing Calibration for an MLC 8000 Subsite Link Converter” in the MLC 8000 Configuration Tool User Guide.
- “Performing Calibration for an MLC 8000 Analog Comparator” in the MLC 8000 Configuration Tool User Guide.
When and where to use:
This process covers optimization/calibration for conventional analog IP-based simulcast systems.Procedure:
- Set up the test equipment. See Setting up Test Equipment.
- Perform coarse level adjustment for the transmit audio path. See Performing Coarse Level Adjustment (Transmit Audio Path 1 kHz).
- Perform a coarse level PL/DPL adjustment. See Performing Coarse Level PL/DPL Adjustment (Conventional Systems).
- Perform fine amplitude adjustment. See Performing Fine Amplitude Adjustment.
- Perform a final system level check. See Performing a Final System-Level Check.
Optimization/calibration is complete.